PRODUCTS
Back to listsTester for Protective Clothing against Flame 【end of manufacture】
Model
FPT-30A
Reference standard
ISO 9151 standard
Use/Good point
This apparatus is used the assessment of thermal behavior of materials and material assemblies when the clothing for protection against heat and fire exposed to a source of flames (predominantly convective heat).
A horizontally oriented specimens, partially restrained from moving is subjected to the flame from a gas burner placed beneath it and the amount of heat passing through the specimen is measured by means of a small copper calorimeter placed behind.
The time/temperature relationship of the calorimeter obtained during the test is used to determine the exposure time required to cause pain and second degree burn or blister in accordance with specified burn criteria.
Remarks
Stand for cooling fan ・・・1 set
Calorimeter ・・・2 pcs
Support frame for specimen ・・・2 pcs
Calorimeter restraining plate ・・・2 pcs
Cleaning solvent for calorimeter ・・・1 bottle (Acetone)
Printer paper ・・・3 pcs
Spec
FPT-30A | |
---|---|
Gas burner | Meckel burner (φ38㎜) |
Thermocouple | Copper wire / constantan (type : T) , Wire diameter : 0.25㎜ |
Calorimeter | Diameter : 40㎜ , Thickness : 1.6㎜ , Purity : 99.55% |
Spacer plate | Dimension : 149×149×6t㎜ (Diameter of center-hole : 90㎜) Weight : 264±13 g |
Gas | Propane gas (With the reducing valve and flow meter on the control panel) |
Shutter for flame | Hand operate (Pull-out system to front and rear) |
Heat flux of gas | 80kW/m2±5% (The line of success or failure indicate on the LCD display) |
Ignition time | 10 seconds (at regulation gas flow) ~30 seconds (at your option in test) |
Checking method for heat flux | Graph (Temp./time) indicate on the LCD display |
Test results | Indication on the display and print-out |
Power source | AC100V 50/60Hz |
Dimensions | W710×D320×H525㎜ approx. |
Weight | 80㎏ approx. |