PRODUCTS

Back to lists

Light Blocking Effect Tester

Model

LBC-10A

Reference standard

JIS L-1055 standard

Use/Good point

This tester is developed to appraise the curtains light blocking effect under the constant condition, based on method A of the JIS L-1055 (lightmeter method).
Turn on a light source (a Halogen electric bulb) and expose light to the light meter photoreceptor. Measure the illumination penetrated the specimen after adjusting light source to a brightness of 10,000 (1x) and obtain shading rate (%) by the following formula.
Shading Rate (%) = (1-I1/I0) × 100
I1 = the illumination with specimen (1x)
I0= the illumination without specimen (1x)

Remarks

【ACCESS.】
Specimen clamping ring (metal)   1pc

Spec

 LBC-10A
Specimen dimension200×200mm
Light sourceHalogen electric bulb JD 100V 500W
Lightmeter~100 000(1x)
Lightmeter accuracy±5% of full scale
Light source adjustmentVoltage adjuster(secondary voltage 0~130V )
Test box dimension400(W)×400(D)×100(H)mm light extinction black coated fabric
PowerAC 100V 500W
Outside dimension600(W)×500(D)×800(H)mm approx